Novel lateral transfer assist robot decreases the difficulty of transfer in post-stroke hemiparesis patients: a pilot study

KOYAMA, Soichiro
TANABE, Shigeo
OTAKA, Yohei
KATO, Tomoya
FURUZAWA, Shotaro
TATEMOTO, Tsuyoshi
KUMAZAWA, Nobuhiro
YOSHIMUTA, Hiroshi
TORII, Katsuhiko
TSUKADA, Shingo
SAITOH, Eiichi
et al

Publication Date 

2020
6 pp

Purpose

The purpose of this study was to clarify whether the novel lateral transfer assist robot facilitates easier transfers compared with a wheelchair in post-stroke hemiparesis patients.

 

Methods

This cross-sectional study enrolled 20 post-stroke hemiparesis patients, and the task difficulty of transfers was compared between a wheelchair and lateral transfer assist robot. All participants were asked to transfer from either wheelchair or lateral transfer assist robot to a platform table and back. The primary outcome was the transfer score of the Functional Independence Measure. The secondary outcome was the time required for transfer.

 

Results

The transfer score of the Functional Independence Measure was significantly higher with lateral transfer assist robot than with wheelchair (p < .001). The transfer times from these devices to a platform table and back showed no significant differences (to device from platform table: 7.8 s, lateral transfer assist robot vs 7.6 s, wheelchair, p > .05: device to platform table: 7.1 s, lateral transfer assist robot vs 8.0 s, wheelchair, p > .05).

 

Conclusions

Transfer with a lateral transfer assist robot is easier than with wheelchair and facilitates independence in post-stroke hemiparesis patients.